MIL-DTL-19246E
3.5.6 Proof pressure. The connector shall not show any evidence of structural damage
when tested as specified in 4.7.7.
3.5.7 Bailout flow. The connector shall permit a minimum flow of 9 Lpm when tested as
specified in 4.7.8.
3.5.8 Endurance. The connector shall disconnect between 12 and 20 pounds when tested as
specified in 4.7.11.
3.5.9 Drop test. The connector shall operate when tested as specified in 4.7.12.
3.5.10 Salt fog. The connector shall not show any evidence of corrosion when tested as
specified in 4.7.13.
3.5.11 Disassembly and examination. The connector shall not show evidence of
deterioration or wear when tested as specified in 4.7.14.
3.5.12 Ozone-resistance. The elastomer components shall not bloom or show any evidence
of checking, cracking, or other damage when tested as specified in 4.7.15.
3.6 Interchangeability. All parts having the same manufacturer's part number shall be
functionally and dimensionally interchangeable.
3.7 Weight. The weight of the complete connector shall be not greater than 0.20 pound.
3.8 Identification of product. The connectors shall be marked for identification in
accordance with MIL-STD-130.
3.9 Marking. Nomenclature tags shall be inserted in each bag showing title, part number or
other data as required by the procuring activity. In addition, interior and exterior containers shall
be marked in accordance with MIL-STD-129.
3.9.1 Precautionary marking. The following precautionary marking shall appear on each
package:
CAUTION: DO NOT ALLOW CONTAMINANTS OF ANY KIND TO BE USED ON OR
ABOUT THE OXYGEN CONNECTORS.
3.10 Workmanship. The connectors shall be uniform in quality and shall be free from
irregularities, defects, or foreign matter which could affect safety, performance, reliability, or
durability.