TABLE II. Dynamic outlet pressure conditions and requirements (see 3.6.2.1).
Breath volume (liters/breath) |
Flowrate Average Peak (liters/min) |
Breathing frequency (breaths/min) |
Regulator outlet pressure (inch H20) |
1.0 2.0 1.0 2.0 1.0 2.0 1.5 2.0 3.0 1.5 2.0 3.0 1.5 2.0 3.0 |
15.9 50 15.9 50 23.9 75 23.9 75 31.8 100 31.8 100 47.7 150 47.7 150 47.7 150 57.3 180 57.3 180 57.3 180 63.7 200 63.7 200 63.7 200 |
15.9 8.0 23.9 12.0 31.8 15.9 31.8 23.9 15.9 38.2 28.7 19.1 42.5 31.9 21.2 |
0.5 to 1.5 0.5 to 1.5 0.7 to 2.0 0.7 to 2.0 1.0 to 2.5 1.0 to 2.5 1.5 to 3.5 1.5 to 3.5 1.5 to 3.5 2.0 to 4.5 2.0 to 4.5 2.0 to 4.5 2.5 to 5.0 2.5 to 5.0 2.5 to 5.0 |
Note:
1. The regulator shall be tested at minimum inlet pressures of 5,8,15,30 and
80 +2/-0 psig at each of the above conditions, except as follows:
a. At an inlet pressure of 5 psig, flowrates above 75 lpm shall not be required.
b. At an inlet pressure of 8 psig, flowrates above 100 lpm shall not be required.
2. Regulator inlet pressure, peak flow and outlet pressure shall be as defined in Figure 3.
3. Breathing inhalation and exhalation shall be sinusoidal (see 4.7.7.4).
4. The duration of testing shall be at least 1 minute.
20